4.8.2 Universal specimen holder and flat specimen holder

Place the specimen into the specimen holder
(5) and (6) slightly tighten screw (Use special
tool!)

Use the special forceps with large radius to
insert the specimen holder into the cryo speci-
men arm and tighten loosely with screw (3) of
cold bridge (4).

Fig. 23

6

5

Release handwheel lock.

Place the specimen cylinder into the lowest possible position to en-
sure a fast freezing process.

Once the desired temperature has been reached, bring the specimen
cylinder back in the upper limit position.

Lock the handwheel.

Tighten screw (6) in specimen holder (5).

Loosen screw (3, Fig. 22) of cold bridge (4, Fig. 22) from the cryo
specimen arm, orient the specimen holder into the desired cutting
direction.

Tighten screw (3) or cold bridge (4) thoroughly.

Extra Fine Screw Thread! Do not tighten too much!