The selected temperature parameters are dis-
played for about 4 seconds. Then the display
switches back to actual temperature.

‘OPERATION’ control panel

With the LN

2

-STOP button (7), the LN

2

pump can

be stopped at any time, e.g. for refilling the Dewar.

The STOP-Alarm button (8) is to switch off the
beep signal, which sounds either when the filling
level of the Dewar falls below 20% of full or when
the lowest LN

2

level in the cryochamber is

reached.

Button H (9) activates chamber heating.

Fig. 18c

7

8

9

11

10

The flow diagram indicates the LN

2

filling level of Dewar vessel and

cryochamber (10). When the lowest filling level inside the Dewar is
reached, the pump is automatically switched off.

To change the unit of measurement that is displayed (K, °C, °F) switch
on the control unit, immediately press the START button (11) and keep
it pressed down for at least 7 seconds.

With the “+” button of the knife temperature indication (2), situated in
the ‘TEMPERATURE’ control panel, you can change between 1 (K),
2 (°C) and 3 (°F) (Fig.18a, 12).

To exit the selection mode press the H-button (9).