5.2.3 Exchanging the bulb

Before replacing the lamp, switch off the mains switch and pull the
mains plug!

The bulb of the chamber illumination is located
behind a cover on the left side of the cryocham-
ber.

•

Remove the two screws (1) of the cover (2)
with 2.5 screwdriver.

Fig. 40

Fig. 41

•

Carefully pull out the cover (2) together with
the cable attached to it on the inside just
enough to be able to reach the lamp.

•

Remove the lamp (3) and insert a new lamp of
the same specification.

•

Put the cover (2) back in its place and secure
with the screws (1).

3

2

2

1

5.

Cleaning and maintenance