Cleaning and maintenance – Leica Biosystems LN22 User Manual
Page 40
Advertising
40
Instructions for Use V1.2, RevB – 10/2012
5.2.3 Exchanging the bulb
Before replacing the lamp, switch off the mains switch and pull the
mains plug!
The bulb of the chamber illumination is located
behind a cover on the left side of the cryocham-
ber.
•
Remove the two screws (1) of the cover (2)
with 2.5 screwdriver.
Fig. 40
Fig. 41
•
Carefully pull out the cover (2) together with
the cable attached to it on the inside just
enough to be able to reach the lamp.
•
Remove the lamp (3) and insert a new lamp of
the same specification.
•
Put the cover (2) back in its place and secure
with the screws (1).
3
2
2
1
5.
Cleaning and maintenance