5 technical data, Technical data, Instrument components and specifications – Leica Biosystems ST4020 User Manual

Page 16

Advertising
background image

16

Instructions for Use V1.1, RevB – 10/2012

3.5 Technical Data

3.

Instrument Components and Specifications

Dimensions and weights

Dimensions, (W x D x H):

620 mm x 250 mm x 200 mm

Height with Hood open:

430 mm

Dry weight, unpacked:

approx. 16 kg

Instrument with accessories:

approx. 19 kg

Weight incl. packing material:

approx. 21 kg

Performance

Slide processing rate:

Depends on program structure and loading

Loading capacity:

max. 4 slide carriers

Stations:

max. 14

Reagent stations:

max. 14

Washing stations:

max. 3 (then only 11 reagent stations are possible)

Reagent container volume:

50 ml

Agitation:

0 to 3 dips per station for dwell times > 30 sec.

Throughput rate of washing stations: 250 ml/min +/-100 ml, depending on number of stations

Dwell time adjustment:

2 to 300 sec. at each station

Capacity of the slide carrier:

max. 4 slides

Capacity of the exit tank:

max. 4 slide carriers

General

Admissions:

UL, cUL, CE

Nominal voltage:

100 V – 240 V +/- 10% AC

Nominal frequency:

50 to 60 Hz

Power draw:

25 VA

IEC 1010 classification:

Protective class 1
Pollution degree 2
Overvoltage installation category II:

Secondary fuses:

Melting fuses 1.0 A T Ø 5.0 x 20 mm

Operating temperature range:

15 °C to 30 °C

Relative humidity:

20% to 80%, non-condensing

Altitude:

< 6500 ft (Ambient pressure. 29 - 31 inch Hg)

Advertising