Setting up the instrument, 1 location conditions, 2 unpacking the instrument – Leica Biosystems ST4020 User Manual

Page 17: Location conditions, Unpacking the instrument

Advertising
background image

17

Leica ST4020

4.1 Location Conditions

The instrument requires an installation area of approx. 250 x 700 mm.
A clearance of at least 25 cm must be left between the instrument and
the closest wall or other instruments.

If the stainer is to be operated with rinsing stations, fresh water and
waste water connections must be no more than 1.2 m from the ports
on the back of the instrument.

It must be ensured that there is sufficient clear space (approx. 30 cm)
above the instrument to allow unobstructed opening of the optional
Hood.

Room temperature consistently between +10 °C and +30 °C.

Relative air humidity maximum 80%, non-condensing.

Practically vibration-free floor. Avoid vibrations, direct sunlight and
heavy variations in temperature.

4.

Setting up the instrument

The chemicals used in the Leica ST4020 are easily inflammable
and hazardous to health.
The installation site for the Leica ST4020 must be well venti-
lated, and there must be no ignition sources of any kind in the
area.
The instrument must not be operated in areas at risk of
explosion.

4.2 Unpacking the Instrument

First check the shipment for external damages upon arrival.
If it is evident that the shipment was damaged during transport,
please make a claim to the carrier immediately.

Open the packaging.

Remove all foam material.

Take out all accessories and the Instructions for Use.

Check the contents of the delivered pack against the order form to
ensure that all components and accessories are included.