Troubleshooting – Leica Biosystems ST4020 User Manual
Page 42
42
Instructions for Use V1.1, RevB – 10/2012
•
Keypad does not respond to
key presses
•
The keypad or the control
board is faulty.
•
Inform Technical Service.
Possible cause
Issue
Corrective action
•
Water does not flow.
•
Water hose is not
connected.
•
Tap is not open.
•
The valve in the instrument
is faulty (does not open) or
the control board is faulty.
•
Adjust the flow rate for the
rinsing water at the tap as
described in Chap. 5.4.
•
The drain hose must not be
kinked.
•
Check the drain hose for
objects that might block it.
•
Place the drain hose so that
water does not have to rise
above the level of the drain
fitting.
•
Water overflows at the rin-
se stations.
•
Tap is open too far.
•
The drain hose is clogged.
•
The drain hose is routed
incorrectly, gradient
insufficient.
7.
Troubleshooting
•
Connect the water hose to
the stainer and the tap.
•
Turn on the tap.
•
Inform Technical Service.