Keypad does not respond to
key presses

The keypad or the control
board is faulty.

Inform Technical Service.

Possible cause

Issue

Corrective action

Water does not flow.

Water hose is not
connected.

Tap is not open.

The valve in the instrument
is faulty (does not open) or
the control board is faulty.

Adjust the flow rate for the
rinsing water at the tap as
described in Chap. 5.4.

The drain hose must not be
kinked.

Check the drain hose for
objects that might block it.

Place the drain hose so that
water does not have to rise
above the level of the drain
fitting.

Water overflows at the rin-
se stations.

Tap is open too far.

The drain hose is clogged.

The drain hose is routed
incorrectly, gradient
insufficient.

7.

Troubleshooting

Connect the water hose to
the stainer and the tap.

Turn on the tap.

Inform Technical Service.