42

Instructions for Use V1.1, RevB – 10/2012

•

Keypad does not respond to
key presses

•

The keypad or the control
board is faulty.

•

Inform Technical Service.

Possible cause

Issue

Corrective action

•

Water does not flow.

•

Water hose is not
connected.

•

Tap is not open.

•

The valve in the instrument
is faulty (does not open) or
the control board is faulty.

•

Adjust the flow rate for the
rinsing water at the tap as
described in Chap. 5.4.

•

The drain hose must not be
kinked.

•

Check the drain hose for
objects that might block it.

•

Place the drain hose so that
water does not have to rise
above the level of the drain
fitting.

•

Water overflows at the rin-
se stations.

•

Tap is open too far.

•

The drain hose is clogged.

•

The drain hose is routed
incorrectly, gradient
insufficient.

7.

Troubleshooting

•

Connect the water hose to
the stainer and the tap.

•

Turn on the tap.

•

Inform Technical Service.