Instructions for Use V1.1, RevB – 10/2012

Cut the hose roughly in the middle between the Y-connector and the
rinsing station.

Use the extension connector (20) included with the package to attach
an additional hose section (21) of the required length.

To do this, measure the length of the hose again, and if necessary cut
it to exactly the required length.

Push both ends of the hose as far as they will go onto the extension
connector, and then secure each hose with a cable tie (18).

Finally, check again that the rinsing water container is positioned
horizontally on the platform.

4.6 Remedy if Hose is too short

If the hose (21) at one of the stations should be too short, or if a hose has
been cut too short by mistake, the following remedy may help:

18

20

21

Fig. 29

4.

Setting up the instrument