Setting up the instrument – Leica Biosystems ST4020 User Manual

Page 23

Advertising
background image

23

Leica ST4020

4.

Setting up the instrument

Place the rinsing water container back in the desired position on the
platform. Check that the rinsing water container is positioned horizon-
tally on the platform.

If the hose is too long, the rinsing water container will tilt to the right; if
it is too short, it will tilt to the left.

If the rinsing water container is not positioned horizontally on
the platform, it cannot be guaranteed that the stainer will func-
tion properly.

Repeat the steps described above for the other rinsing stations.

Installing the Rinsing Water Containers

(continued)

If only one or two rinsing stations are required,
the hoses that are not used must be sealed with
a plug (17), see Fig. 27.

To do this, twist and push the plug fully into
the end of the hose, and then secure it with
a cable tie (18).

17

18

Fig. 27

If two rinsing stations are to be located immedi-
ately beside one another, the connector (19) for
the left station must be aligned as shown in

Fig. 28

.

19

Fig. 28