4.
Setting up the instrument
Marking 1
Installing the Rinsing Water Containers
(continued)
Marking 2
Lateral surface of the connector
End of the connector
•
Now take the rinsing water
container and the hose out
of the stainer.
•
Align Marking 1 on the end
of the connector and make
a second marking ("Mark-
ing 2") to indicate the lateral
surface of the connector
(Fig. 25).
•
Cut the hose off at Mark-
ing 2.
•
Then push the hose (21)
fully onto the connector by
twisting until the end of the
hose is flush against the
lateral surface of the con-
nector (Fig. 26).
•
If the hose is not pushed
completely onto the con-
nector, it may become
detached as soon as the
water supply is turned on.
Do not stretch tub-
ing to fit on the fit-
ting.
Fig. 25
Fig. 26
Marking 2
21