Operating condition requirements
Introduction
Operating condition requirements
Only use the high temperature furnace under the following conditions.
Technical parameters
P
ARAMETER
S
PECIFICATION
Condition
For indoor application only
Operating Temperature
5 ~ 40
°
C (41 ~ 104
°F
)
Operating Humidity
5-85%, noncondensing
Maximum Altitude
2000 m (6562 ft)
Voltage
Single-phase AC 200 ~ 240 V
Frequency
50 Hz / 60 Hz
Current
15 Amps
M
ODEL
FGW900
Scope of application
C45, C64
Temperature scope (
°
C)
300 ~ 1100(572
°
F to 2097 °F)
External dimension (mm)
Φ310 x 410(12.2 in x 16.1 in)
Dimension of chamber (mm)
Φ100 x 320(3.9 in x 12.6 in)
Controller size (mm)
400 x 420 x 200(15.7 in x 16.5 in x 7.9 in)
Uniform temperature zone length(mm)
100(3.9 in)
Temperature gradient
≤3
°
C(≤37.4
°
F)
Heating method
Three-segment controllable
Temperature fluctuation
≤± 2
°
C(≤35.6
°
F)
Precision of temperature controller
Class 0.3
Time to obtain operating temperature
≤50 min
Weight
25 kg(55.1lb)