High-Temperature Furnace Cable Connections

FGW900 High-temperature Furnace Product Manual

Installation

21

Connection steps:

1.

Connect the thermocouple wire

Aaccording to the label sequence of J8, J9, and J10, insert thermocouple into the
upper, middle, lower jacks of the high temperature furnace side wall and tighten
the fixing screws. Insert the cable that is carried by the thermocouple into the
thermocouple outlet on the rear panel of the temperature controller.

2.

Connect the door switch wire

Insert one end of switch lead with plug into J7 jack of the temperature controller,
then connect the other end to the limit switch contact on the furnace bottom.
When installing, the furnace door switch lead has been connected together with
the limit switch. Only need to insert the other end to J7 of the temperature
controller.

3.

Connect the output power cable between the furnace and the temperature
controller.

There are two power cables that connect the furnace and the temperature
controller. One is OUTPUTL, the other is OUTPUT N. The two cables have
already been connected together on the furnace. When installing, simply need to
follow the labels to insert one end with plug into the corresponding sockets on
the rear panel of the temperature controller and then tighten the fastens.

Note

Do not insert or unplug this connectors group in the heating condition to
avoid personal electric shock accidents.