Routine Maintenance

FGW900 High-temperature Furnace Product Manual

Routine Maintenance

35

When using the furnace, try not to make contact with the inner cavity of the
furnace during the process of installing a test sample to avoid causing
damage to the furnace cavity.

Do not open the furnace immediately after a test is completed to avoid
serious injury and prevent damage to the furnace cavity.

Clean the furnace cavity in a timely manner with a soft brush once a test is
completed in order to prolong the service life of the furnace.

If the furnace is not used for a long period, preheat the furnace before using
it again. The preheating method is as follows:

1. Increase

the

temperature to 200°C and maintain that temperature for a

least four hours.

2. Increase the temperature to 400°C and maintain that temperature for a
least four hours.

3. Increase the temperature to 600°C and maintain that temperature for a
least four hours.

Treatment of general fault

F

AULT

CONDITION

R

EASON

S

OLUTION

SV window flashes
and displays “S.br”

The input measuring signal
exceeds the measuring
range

1. Check whether the
specifications of the sensor are
properly defined

2. Check whether the signal
input is broken

Heatup speed is
slower than original
speed

The parameter are not
properly defined

Check whether the parameters
of the temperature controller are
properly defined

The Temperature
controller can not start
the furnace

The Furnace is not closed
properly or there is
something wrong with the
Furnace door switch or cable

1.Check the furnace of see if it is
closed properly

2.Check the door switch cable
and the connections to the
furnace door switch and the back
of the temperature controller

Temperature fails to
steady at setting
temperature

Control parameter is
imperfect or lost

Start self-adjustment to find out
proper control parameter