MTS FGW900 High-temperature Furnace User Manual

Page 24

Advertising
background image

High-Temperature Furnace Cable Connections

FGW900 High-temperature Furnace Product Manual

Installation

24

Connecting the Main

Power

The input voltage of the MTS FGW900 high temperature Furnace is single phase
200-240 V, 50/60 Hz. Use the cable that is provide for power input, and connect
to the customer electrical box and disconnect.

Note

Local electrical codes supercede any information found here. Electrical
connections must be made by qualified personnel and is their
responsibility for using the proper power disconnect that conforms to
local codes and regulations when connecting the machine to the building
main power.

Electrical disconnect

The customer is responsible for providing an electrical power disconnect that is
easy to operate and easy to reach. It must also meet IEC 60947-1 and IEC 60947-
3 standards. Recommended circuit breaks would be ones that are of the thermal
magnetic type with characteristics suitable for large inductive loads(D-type trip
characteristic). If fuses are used it is recommended that they are of the time delay
type with dual elements. These recommendations should be followed to avoid
nuisance tripping.