Operation of High Temperature Furnace

FGW900 High-temperature Furnace Product Manual

Operation

26

Operation of High Temperature Furnace

Before starting test,you need assemble high temperature grip, open high
temperature furnace and revolve it into test space, assemble high temperature
grip into furnace body and seal pull rod hole according to requirement.

The high temperature furnace belongs to split flange high temperature furnace, of
which opening/closing is realized via lock on external wall of case shell.

The bottom of the high temperature furnace is equipped with a limit protection
switch, when the high temperature furnace is opened, this limit protection switch
will cut off the heating power, and can not self-recovery, only when to again
press the ENABLE button can be connected to the heating circuit once the door
is closed again.

If you use a high-temperature extensometer (supplied separately), its extension
guide rod is inserted into the furnace via the slot for the extensometer in front of
the furnace and contacts the testing sample. For information about the installation
and usage of the high-temperature extensometer, see the high-temperature
extensometer user manual.

Furnace Handle

Extensometer Channel

Furnace Lock

Thermocouple Hole