2. Check the inflation pressures on each airmount. Bleed any airmount whose pressure is above 0.55 MPa
(80 psi).
3. Check the airmounts’ inflated heights. They should be between 57–63 mm (2.25–2.5 in).
If the load unit is level and the airmount heights correct, you are done.
If not, continue this procedure.
Checking Airmount Inflated Heights
Description
Item
5 mm (0.20 in)
1
0.41 MPa (60 psi)
2
57–63 mm (2.25–2.5 in)
3
4. Check the airmounts’ inflated heights. They should be between 83–86 mm (3.25–3.5 in).
If the load unit is level and the airmount heights correct, you are done.
If not, continue this procedure.
MTS Landmark™ Floor Standing Load Units Product Information 85