Adjusting the Hydraulic Locks
Description
Item
Tighten
1
Bottom Out Position
2
Loosen, then hand tighten.
3
7. Loosen each of the hydraulic lock cap screws 1/4 turn.
Loosening the Hydraulic Lock Cap Screws
Description
Item
Loosen 1/4 turn.
1
8. Turn on electrical power at the test controller if you have not already done so.
9. Reset any active interlocks at the test controller.
10. Turn on high hydraulic pressure.
11. Return the Lock Control to the lock crosshead position to pressurize the hydraulic locks.