Crosshead Fastener Torque Values
Description
Item
136 N·m (100 lbf·ft)
1
271 N·m (200 lbf·ft)
2
271 N·m (200 lbf·ft)
3
4. Turn off hydraulic pressure.
5. Use the Lock Control to unlock the crosshead position to remove pressure from the hydraulic locks. Wait
two minutes for the pressure in the locks to drop to zero before going on to the next step.
6. Tighten each lock’s cap screw until its piston bottoms out. Then loosen and hand-tighten each cap screw.
88 MTS Landmark™ Floor Standing Load Units Product Information