Series 661 force transducer – MTS Landmark Floor Standing User Manual

Page 35

Advertising
background image

Series 661 Force Transducer

The force transducer commonly used with the Model 370 Load Unit is a Series 661 Force Transducer. There
is no manual for the force transducer. The following table provides the specifications for the force transducers.

Specification

Parameter

15 V DC

Maximum excitation voltage

350 ¾

Bridge resistance

1.0% of full scale torsional to load

Maximum crosstalk

0.08% of full scale (250 N–2.5 kN)

Hysteresis

0.05% of full scale (5 kN–50 kN)

0.15% of full scale (100 kN–500 kN)

0.20% of full scale (1000 kN)

0.08% of full scale

Nonlinearity

0.15% of full scale for Models 661.22/.23/.31

0.004% of reading/°C (0.002%/°F)

Temperature

-54°C to +121°C (-65 °F to +250 °F)

Usable range

+21°C to +77°C (+70 °F to +170 °F)

Compensated range

0.0036% of full scale/°C (0.0020% or full scale/°F)

Sensitivity

2 mV/V at full-scale load

Output

JT,RCPT,13PIN #10,BOX MNT,.719MTPAT

Connector

PT02ER-10-6P

TEDS

non-TEDS

Weight

Thread Size

Load Capacity

Model

9.75 kg (21.5 lb)

M27 x 2.0 mm x 31.7 mm

25 kN (5.5 kip)

661.20-01

9.75 kg (21.5 lb)

M27 x 2.0 mm x 31.7 mm

50 kN (11 kip)

661.20-02

9.75 kg (21.5 lb)

M27 x 2.0 mm x 31.7 mm

100 kN (22 kip)

661.20-03

13.2 kg (29 lb)

M36 x 2.0 mm

250 kN (55 kip)

661.22-01

16 kg (35.3 lb)

M52 x 2.0 mm x 48.3 mm

500 kN (110 kip)

661.23-01

MTS Landmark™ Floor Standing Load Units Product Information 35

Introduction

Advertising
This manual is related to the following products: