Special transducer considerations, Control module – MTS Landmark Floor Standing User Manual

Page 58

Advertising
background image

Special Transducer Considerations

Special considerations must be followed when using low force transducers in a high force system. Small force
transducers are commonly used in larger systems. You need to be aware that the load unit can produce
forces that exceed the rating of the low force transducer.

Caution:

Low force transducers in a high force system can be exposed to excessive forces.

Applying forces that exceed the rated capacity of the force transducer can damage the transducer.

Ensure that all test commands are within the force transducer’s rating when it is installed. Your
warranty does not cover damage caused by a high force actuator.

MTS recommends the following:

Position the crosshead so that fixtures cannot touch each other when the actuator is fully extended. This
may not be feasible at all times, but it should be practiced whenever possible.

Use force limits. Force limits can help in some instances, but might not stop the actuator fast enough in
all situations.

Ensure that you have selected the correct calibration for the transducer being used. Also ensure that the
system tuning has been optimized for the transducer being used.

Control Module

The controls for the Series 370 Load Unit are located on a module mounted to the front of the load unit. The
followingcontrol panel is shown with all available options. The control panel on your load unit might not have
all of these controls depending on the specific configuration.

58 MTS Landmark™ Floor Standing Load Units Product Information

Operation

Advertising
This manual is related to the following products: