Set travel limits – MTS Criterion Series 40 User Manual
Page 64
The status window appears. Tripped interlocks are indicated with red boxes. (The two interlock conditions
shown below are normal after startup.)
2. Correct any interlock conditions and then reset the interlock.
Click the Reset button in the Controller panel.
Set Travel Limits
Setting the travel limit stops to positions just outside expected test crosshead travel helps prevent equipment
damage and injury. The interlock will set and the crosshead will stop whenever the crosshead reaches a
travel limit stop.
MTS recommends setting limits twice. First, before installing the specimen, set limits just outside the range
of travel necessary to safely install the specimen. Second, before running the test, set limits to just outside
the expected range of travel for the test you are running.
For details on what to do if a limit is tripped, see
Recover from a Tripped Travel Limit
(p. 66)
64 | MTS Criterion
™
Series 40
Running the Example Tension Test