Set up the load meter for specimen installation – MTS Criterion Series 40 User Manual
Page 70
Advertising
c) If a travel limit stop prevents movement, see
Recover from a Tripped Travel Limit
(p. 66).
Set Up the Load Meter for Specimen Installation
This procedure is performed to ensure that the forces imparted on the specimen during installation are
acceptable and have not damaged the specimen.
1. If necessary, expand the meter area at the bottom of the window by clicking the small expand tab in
the lower center of the window.
2. Select the desired units.
70 | MTS Criterion
™
Series 40
Running the Example Tension Test
Advertising