Set up the load meter for specimen installation – MTS Criterion Series 40 User Manual

Page 70

Advertising
background image

c) If a travel limit stop prevents movement, see

Recover from a Tripped Travel Limit

(p. 66).

Set Up the Load Meter for Specimen Installation

This procedure is performed to ensure that the forces imparted on the specimen during installation are
acceptable and have not damaged the specimen.

1. If necessary, expand the meter area at the bottom of the window by clicking the small expand tab in

the lower center of the window.

2. Select the desired units.

70 | MTS Criterion

Series 40

Running the Example Tension Test

Advertising