Unplug lines – MTS Triaxial Cell Models 656-06 User Manual

Page 79

Advertising
background image

Description

Item

Tighten to 20 lb•ft (28 N•m)

2

Caution

:

Install the insert so the side with three holes faces out. If this side faces in, the insert’s
O-ring may leak.

6. Install the O-ring on the force transducer’s insert. Then use a nylon mallet to tap the insert in place.

Install the Insert

Description

Item

Holes must face out.

1

O-Ring

2

Tap in place.

3

7. Connect the force transducer’s cable to its feedthrough in the platen.

Unplug Lines

1. Remove the plug on the lower vent fitting. Install a vent hose on this fitting. Route the hose to a clean

5 gallon (20 litre) pail.

If you are going to use pore pressure in your test, unplug the two pore pressure ports.

Triaxial Cell Product Information | 79