MTS Triaxial Cell Models 656-06 User Manual

Page 68

Advertising
background image

Remove the Force Transducer

Note: Inspect and clean the shoulder bolt’s threads. Then lightly lubricate them with clean
confining fluid. Also lubricate all dry O-rings with clean confining fluid.

5. Install the O-rings as shown in this figure. Tighten the shoulder bolt to 20 lb•ft (28 N•m).

Install the Other Force Transducer

Description

Item

O-Ring

1

Tighten to 20 lb•ft (28 N•m)

2

Caution

:

Install the insert so the side with three holes faces out. If this side faces in, the insert’s
O-ring may leak.

6. Install the O-ring on the force transducer’s insert. Then use a nylon mallet to tap the insert in place.

68 | Triaxial Cell Product Information

Operation

Advertising
This manual is related to the following products: