Transport, scope of delivery, installation, 10 cooling, Warning – Retsch Emax User Manual

Page 20

Advertising
background image

Transport, scope of delivery, installation

20

Fig. 7: Removing the transport aid

4.10 Cooling

WARNING

Risk of a fatal electric shock

-

An electric shock can cause injuries in the form of burns and cardiac

arrhythmia, respiratory arrest or cardiac arrest.

Do not clean the blender under running water. Use only a cloth dampened

with water.

• Disconnect the power supply plug before cleaning the blender.

WARNING

W0008

Risk of death through power surge
Mains plug not completely plugged in

Water can enter the IEC socket if the IEC is not completely plugged in. An
electric shock can lead to burns and arrhythmia or to respiratory arrest as
well as cardiac arrest.

Only operate the device with the IEC fully plugged in.

4.10.1 Cooling agent

There is a cooling agent tank at the back which must be filled with water before
putting into operation. Only clean, lime-free water is permitted as cooling agent.

Unscrew the lid (TD) to fill the water.

Start the device up for 2 minutes at a speed of 300 revolutions per minute
(with two grinding jars).

This activates the internal pump and the cooling agent is distributed in the
cooling system.

Then close the tank again.

TS