Cleaning, wear and service
56
8.4.1
Replacing the cooling water
8.4.1.1
Removing the cooling water
Fig. 36: Replacing the cooling agent
1.
Place a collecting vessel (AG) for the old cooling water under the cooling
agent tank (T).
Fig. 37: Removing the lid
2.
Unscrew the lid (TD).
3.
Remove the sealing screw (DS) to drain the water.
4.
Start the device up for 1 minute at a speed of 300 revolutions per minute
(with two grinding jars).
–
This activates the internal pump (U) and the remaining cooling agent is
pumped out.