Operating the machine

33

Pull out the locking pin (D) to release from the groove (DN).

Twist the locking pin 90 degrees to release it permanently.

Turn the twist grip (C) to open the grinding jar support.

Only remove the grinding jar with closed lid.

Only open the grinding jar in a safe position (e.g. extraction device).

Only open the grinding jar once it has cooled down.

Fig. 20: Removing the grinding jar

5.5.2

Closing

With this grinding device, a very large amount of energy enters the sample
material.

Therefore make sure that the grinding jar support has been closed properly.

Before grinding check the lock on the grinding jar support (V).

Fig. 21: Locking the grinding jar support

The twist grip (C) of the grinding jar support (G) is secured by locking pin (D) to
prevent inadvertent opening.

Twist the locking pin (D) until it engages in the groove (DN).

G

DN

C

D

V