RKI Instruments Beacon 110 User Manual

Page 11

Advertising
background image

Beacon 110 Operator’s Manual

7

Figure 2: Control PCB Component Location

LCD Display

The LCD display is located at the top of the control PCB. It indicates the current gas
reading and displays messages and parameters in the Beacon 110’s operating modes.

Contrast Potentiometer

The contrast potentiometer is located to the right of the LCD display. It is used to adjust
the contrast of the LCD. If the background of the LCD appears so dark that the characters
are not visible or if the characters are too dim, turn the adjustment screw on the
potentiometer clockwise or counterclockwise until the desired visibility is obtained.

Control Buttons

The Beacon 110 includes three push button switches that allow you to enter the Beacon
110’s operating modes, navigate through the modes, update settings, and save changes
to the settings. The push button switches are located along the bottom edge of the control
PCB (see Figure 2). The UP/YES button is on the left, the DOWN/NO button is in the
middle, and the ENTER button is on the right.

Table 4: Beacon 110 Control Button Functions

Switch

Function

UP/YES

Saves settings

Changes the displayed setting

Enters Calibration Mode

Enters Input Mode (press with DOWN/NO button)

Enters Configuration Mode (press with ENTER button)

DOWN/NO

Cancels setting changes

Changes the displayed setting

Enters Input Mode (press with UP/YES button)

Displays the Information Screen

UP/YES DOWN/NO ENTER

F

A1

A2

Control Switches

Fail LED

Alarm 1 LED

Alarm 2 LED

LCD Display

Contrast
Potentiometer