Beacon 110 Operator’s Manual

31

6.

If you changed the input type above, you must select a gas type. Proceed to step 7.

If you did not change the input type, you can decide if you want to select the gas type.
If you do not want to select the gas type, press and release the DOWN/NO button.
The display will indicate NO CHANGE and the Beacon 110 will return to normal
operation with the previous setup. If you do want to select the gas type, press the UP/
YES button and the display will indicate GasType? on the top display line and the gas
type choice on the bottom display line.

7.

Use the DOWN/NO button to scroll through the choices of gas types. When you come
to the desired choice, press the UP/YES button to select it.

If you selected one of the defined gas types such as OXYGEN or CO, the display will
ask SAVE IT? YES/NO. Proceed to step 8.

If you selected the OTHER gas type, the display will indicate GAS NAME on the top
display line and bottom line will be blank with the cursor flashing in the far left.
Proceed to step 9.

8.

If you did not change the input type above and want to discard the gas type selection,
press and release the DOWN/NO button. The display will ask DO OVER? YES/NO.
Press the DOWN/NO button. The display will ask ABORT? YES/NO. Press the UP/
YES button. The display will indicate Aborting Gas Type, then NO CHANGE and the
Beacon 110 will return to normal operation.

If you want to save the gas type selection, press and release the UP/YES button. The
display will indicate Gas Type Updated, show the target gas and the detection range
for a few seconds, then the Beacon 110 will enter Configuration Mode so that the
parameter settings in Configuration Mode can be verified or changed.

9.

Perform the following steps to define the gas name, detection range, and units if you
selected the OTHER gas type.

With GAS NAME on the top display line and the cursor flashing in the far left on
the bottom display line, use the UP/YES and DOWN/NO buttons to select a
character for the first character of the gas name and press ENTER to save it.
The cursor will move to the next position. Repeat this process until the last
character is saved.

The display will ask SAVE IT? YES/NO. Press and release the UP/YES button.
The display will indicate RANGE . . . on the top display line prompting you to
enter a full scale value and the bottom line will be blank with the cursor flashing
on the far left.

Use the UP/YES and DOWN/NO buttons to enter a number for the first digit of
the full scale value and press ENTER to save it. The cursor will move to the next
position. Repeat this process until you have entered the full scale value, for
example 30.0, leaving extra positions blank before pressing ENTER.

When you have pressed ENTER for the last time, the display will ask SAVE IT?
YES/NO
. Press and release the UP/YES button. The display will indicate
UNITS? on the top line and the units choice on the bottom line of the display.

Use the DOWN/NO button to scroll through the choices until you find the
desired units. (If you select the OTHER choice, perform the same procedure
described above for entering the gas name and range.)

Press the UP/YES button to save the units. The Beacon 110 will proceed to
Configuration Mode.