Appendix a: control button quick reference guide – RKI Instruments Beacon 110 User Manual

Page 46

Advertising
background image

42

Beacon 110 Operator’s Manual

Appendix A: Control Button Quick Reference Guide

The Beacon 110’s control buttons and reset switch allow access to operational modes,
resetting of alarms, and display of the Information Screen. Table 11 shows which button
combinations perform these functions and which parameters are available for adjustment
while in the operational modes. While in these modes, display prompts showing a “?”
require you to respond by pressing either the UP/YES (for yes) or DOWN/NO (for no)
button. To change parameter settings, use the UP/YES and DOWN/NO buttons to get the
desired setting, then press and release the ENTER button to accept the setting.

Table 11: Control Button Quick Reference Chart

Mode Entered or

Function Performed

Button

Combination

Adjustable or Viewable Parameters

Alarm Reset

Press and Release
the Reset Switch

n/a

Min/Max

Press and Hold the
Reset Switch for 3
seconds

Minimum and maximum readings can be viewed
and reset.

Show Information Screen

Press and Hold
DOWN/NO

Operating DC Voltage. For AC powered units,
this is the DC voltage from the power supply.

Beacon 110 Version/Revision Information

Calibration Mode

Press and Hold
UP/YES for 5 sec-
onds

Fresh Air Adjust

Span (or Zero for Oxygen) Adjust with
Calibration Gas

Configuration Mode

Press and Hold
UP/YES & ENTER
for 5 seconds

Alarm Levels

Increasing or Decreasing Alarms

Normally Energized or Normally De-energized
Relays

Latching or Self-resetting Alarms

Strobe operation (if installed)

Alarm On Delay

Alarm Off Delay (if self-resetting)

Zero Suppression

Noise Filter

Calibration Time-out Period

Input Mode

Press and Hold
UP/YES and
DOWN/NO for 5
seconds

Define Detector Head Input

Advertising