RKI Instruments Beacon 110 User Manual

Page 31

Advertising
background image

Beacon 110 Operator’s Manual

27

Table 7: Configuration Parameters

Parameter

(Factory Set Value)

Description

ALARM-1 (level)
See the Beacon 110 Detector Head
Specification Sheet for the detector
head installed

The gas reading at which the Beacon 110 initiates an alarm 1 condition.

ALARM-1 (activation)
(Decrease for oxygen, Increase for all
other types)

Indicates if the alarm 1 circuit is activated by gas readings increasing
(Increase) or decreasing (Decrease) to the ALARM-1 Level.

ALARM-1 (relay action)
(N. DE-EN)

If set as N. DE-EN, the alarm 1 relay is de-energized in normal operation
and energizes when an alarm 1 condition is initiated.
If set as N. EN, the alarm 1 relay is energized in normal operation and
de-energizes when an alarm 1 condition is initiated.

ALARM-1 (relay reset)
(SELF-RST for oxygen, LATCH for all
other channel types)

If set as LATCH, you must press the reset switch to reset the alarm 1 cir-
cuit after the alarm 1 condition passes.
If set as SELF-RST, the Beacon 110 automatically resets the alarm 1 cir-
cuit after the alarm 1 condition passes.

A1Strobe
(None if no strobe is installed, Non
Reset if a strobe is installed)

If set as None, there is no strobe operation.
If set as NonReset, it is not possible to turn off the strobe with the reset
switch while the Beacon 110 is in an alarm 1 condition.
If set as CanReset, the strobe can be turned off while the Beacon 110 is
in an alarm 1 condition by pressing the reset switch.

A1 OnDy (alarm 1 on delay)
(1 secs)

The amount of time the Beacon 110 delays activation of the alarm 1 cir-
cuit once an alarm 1 condition is initiated.

A1 OffDy (alarm 1 off delay)
(0 sec.)

The amount of time the Beacon 110 delays turning off the alarm 1 circuit
once an alarm 1 condition passes. This parameter appears only if the
alarm 1 relay reset setting is set to SELF--RST.

ALARM-2 (level)
See the Beacon 110 Detector Head
Specification Sheet for the detector
head installed

The gas reading at which the Beacon 110 initiates an alarm 2 condition.

ALARM-2 (activation)
(Increase)

Indicates if the alarm 2 circuit is activated by gas readings increasing
(Increase) or decreasing (Decrease) to the ALARM-1 Level.

ALARM-2 (relay action)
(N. DE-EN)

If set as N. DE-EN, the alarm 2 relay is de-energized in normal operation
and energizes when an alarm 2 condition is initiated.
If set as N. EN, the alarm 2 relay is energized in normal operation and
de-energizes when an alarm 2 condition is initiated.

ALARM-2 (relay reset)
(SELF-RST for oxygen, LATCH for all
other channel types)

If set as LATCH, you must press the ENTER button to reset the alarm 2
circuit after the alarm 2 condition passes.
If set as SELF-RST, the Beacon 110 automatically resets the alarm 2 cir-
cuit after the alarm 2 condition passes.