RKI Instruments Beacon 110 Rig Monitor User Manual
Page 28

23
Beacon 110 Rig Monitor Operator’s Manual
Table 6: Configuration Parameters
Parameter
(Factory Set Value)
Description
ALARM-1
(10 ppm)
The gas reading at which the Rig Monitor initiates an alarm 1 condition.
ALARM-1 (activation)
(Increase)
Indicates if the alarm 1 circuit is activated by gas readings increasing
(Increase) or decreasing (Decrease) to the ALARM-1 Level.
•
Note: Changing this parameter will affect the operation of the factory
installed strobe/horn. Do not change this parameter from the factory
setting.
ALARM-1 (relay action)
(N. DE-EN)
If set as N. DE-EN, the alarm 1 relay is de-energized in normal operation
and energizes when an alarm 1 condition is initiated.
If set as N. EN, the alarm 1 relay is energized in normal operation and
de-energizes when an alarm 1 condition is initiated.
•
Note: Changing this parameter will affect the operation of the factory
installed strobe/horn. Do not change this parameter from the factory
setting.
ALARM-1 (relay reset)
(LATCH)
If set as LATCH, you must press the reset switch to reset the alarm 1 cir-
cuit after the alarm 1 condition passes.
If set as SELF-RST, the Rig Monitor automatically resets the alarm 1 cir-
cuit after the alarm 1 condition passes.
A1Strobe
The strobe controlled by this parameter is not installed in the Rig Monitor.
A1 OnDy (alarm 1 on delay)
(1 secs)
The amount of time the Rig Monitor delays activation of the alarm 1 cir-
cuit once an alarm 1 condition is initiated.
A1 OffDy (alarm 1 off delay)
(0 sec.)
The amount of time the Rig Monitor delays turning off the alarm 1 circuit
once an alarm 1 condition passes. This parameter appears only if the
alarm 1 relay reset setting is set to SELF--RST.
ALARM-2
(50 ppm)
The gas reading at which the Rig Monitor initiates an alarm 2 condition.
ALARM-2 (activation)
(Increase)
Indicates if the alarm 2 circuit is activated by gas readings increasing
(Increase) or decreasing (Decrease) to the ALARM-1 Level.
ALARM-2 (relay action)
(N. DE-EN)
If set as N. DE-EN, the alarm 2 relay is de-energized in normal operation
and energizes when an alarm 2 condition is initiated.
If set as N. EN, the alarm 2 relay is energized in normal operation and
de-energizes when an alarm 2 condition is initiated.
ALARM-2 (relay reset)
(LATCH)
If set as LATCH, you must press the ENTER button to reset the alarm 2
circuit after the alarm 2 condition passes.
If set as SELF-RST, the Rig Monitor automatically resets the alarm 2 cir-
cuit after the alarm 2 condition passes.
A2Strobe
The strobe controlled by this parameter is not installed in the Rig Monitor.
A2 OnDy (alarm 2 on delay)
(1 secs)
The amount of time the Rig Monitor delays activation of the alarm 2 cir-
cuit once an alarm 2 condition is initiated.
A2 OffDy (alarm 2 off delay)
(0 sec.)
The amount of time the Rig Monitor delays turning off the alarm 2 circuit
once an alarm 2 condition passes. This parameter appears only if the
alarm 2 relay reset setting is set to SELF-RST.