Beacon 110 Rig Monitor Operator’s Manual

34

Buzzer Not
Working

The buzzer does not
sound an audible
alarm during alarm
conditions.

The buzzer sounds
weak or broken.

The buzzer is
disconnected or
misconnected.

The buzzer is
malfunctioning.

1. Verify that the wiring to the ALARM

BUZZER +

& - terminals of the

Controller Terminal Strip is correct
and secure.

2. If the buzzer difficulties continue,

contact RKI for further instruction.

Reset Switch
Not Working

The buzzer does not
silence when you
press the reset
switch.

The alarm 1 or
alarm 2 circuit does
not reset when you
press the reset
switch after the
alarm condition
passes.

The reset switch is
disconnected or
misconnected.

The reset switch is
malfunctioning.

1. Verify that the wiring to the ALARM

RESET

(2) terminals of the Alarm/

Controller Terminal Strip is correct
and secure.

2. If the reset switch difficulties continue,

contact RKI for further instruction.

Strobe/Horn
Not Working

The strobe/horn
does not activate
when an alarm 1
condition occurs.

The strobe/horn is
malfunctioning.

The strobe/horn
wiring is disconnected
or misconnected.

1. Verify that the strobe/horn is wired

properly to the alarm terminal strip.
The red wire goes to the + terminal
and the black wire goes to the -
terminal.

2. If the strobe/horn difficulties continue,

contact RKI for further instruction.

Fail Condition

Rig Monitor indicates a
fail condition.

• The H

2

S detector

wiring is disconnected
or misconnected.

• The plug-in H

2

S

sensor is not properly
plugged into the
sockets in the detector
housing body.

• The detector’s zero

reading is low enough
to cause a fail
condition.

• The detector is

malfunctioning.

1. Verify that the H

2

S detector wiring is

correct and secure.

2. Confirm that the plug-in H

2

S sensor is

installed properly.

3. Perform a zero adjustment. A full

calibration is recommended.

4. If the fail condition continues, replace

the plug-in sensor as described later
in this section.

5. If the fail condition continues, contact

RKI for further instruction.

Table 7: Troubleshooting the Beacon 110 Rig Monitor (Continued)