Beacon 110 Rig Monitor Operator’s Manual

Appendix A: Control Button Quick Reference Guide

The Beacon 110 Rig Monitor’s control buttons and reset switch allow access to operational modes,
resetting of alarms, and display of the Information Screen. Table 9 shows which button
combinations perform these functions and which parameters are available for adjustment while in
the operational modes. While in these modes, display prompts showing a “?” require you to
respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button. To change
parameter settings, use the UP/YES and DOWN/NO buttons to get the desired setting, then press
and release the ENTER button to accept the setting.

Table 9: Control Button Quick Reference Chart

Mode Entered or

Function Performed

Button

Combination

Adjustable or Viewable Parameters

Alarm Reset

Press and Release
the Reset Switch

n/a

Min/Max

Press and Hold the
Reset Switch for 3
seconds

Minimum and maximum readings can be viewed
and reset.

Show Information Screen

Press and Hold
DOWN/NO

Operating DC Voltage.

Beacon 110 Rig Monitor Version/Revision
Information

Calibration Mode

Press and Hold
UP/YES for 5
seconds

Fresh Air Adjust

Span Adjust with Calibration Gas

Configuration Mode

Press and Hold
UP/YES & ENTER
for 5 seconds

Alarm Levels

Increasing or Decreasing Alarms

Normally Energized or Normally De-energized
Relays

Latching or Self-resetting Alarms

Strobe operation (if installed)

Alarm On Delay

Alarm Off Delay (if self-resetting)

Zero Suppression

Noise Filter

Calibration Time-out Period