RKI Instruments Beacon 110 Rig Monitor User Manual

Page 40

Advertising
background image

35

Beacon 110 Rig Monitor Operator’s Manual

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Detector responds

slowly or does not
respond during
response test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Detector requires

frequent calibration.

Note:

Under “normal”

circumstances, the
detector requires
calibration once every 3
months.

Some applications may

require a more frequent
calibration schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The flame arrestor in

the detector housing
cap is wet or clogged
with dirt or other
particulates.

• The calibration gas is

not an appropriate
concentration.

• The detector is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Check the detector housing cap to

determine if the flame arrestor is wet
or dirty. Clean if necessary.

3. Verify that the calibration gas

concentration is appropriate for the
detector. Zero emission air (0 ppm
H

2

S) is normally used for a zero

adjustment if the environment is
suspect and 25 PPM H

2

S in nitrogen

is normally used for a response
adjustment.

4. If the calibration/response difficulties

continue, replace the plug-in sensor
as described later in this section.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Table 7: Troubleshooting the Beacon 110 Rig Monitor (Continued)

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising