Digester Gas Monitor Operator’s Manual

Overview • 5

Chapter 2: Description

Overview

This chapter describes the Digester Gas Monitor’s external and internal components.

External Description

This section describes the housing and all external components of the Digester Gas
Monitor. For the purposes of this description, the housing door is considered the front of
the monitor.

Figure 1: External Component Location

Housing

The Digester Gas Monitor’s fiberglass housing is weather- and corrosion-resistant. It is
suitable for installation where general purpose equipment is in use. The housing door is
hinged on the left side and is secured by two latches on the right side. The LCD display,
status LEDs, and flowmeters are visible through the window in the housing door. Four
mounting feet are attached to the back of the housing (one at each corner). The mounting
feet allow you to install the housing to a vertical surface. These mounting feet are not
installed as shipped from the factory (see “Mounting the Digester Gas Monitor” on page 22
for instructions to install the mounting feet). Three conduit hubs on the bottom of the
housing are for external wiring connections.

Reset Switch

The reset switch is on the bottom right of the housing in front of the buzzer. The reset
switch serves four functions:

Resets the alarm circuits for “latched” alarms after an alarm 1, alarm 2, or alarm 3
condition passes.

You can set each channel for latched or self-resetting alarms in the Configuration
Menu. See “Viewing and Changing Channel Parameters” on page 51 for more
information.

Buzzer

Fresh Air Inlet

Reset Switch,
Push Button

Dryer

Exhaust

Conduit Hub, 3/4"

Aspirator/Sample, Exhaust

Compressed

Air Inlet

Blow Back Exhaust

Calibration Gas Inlet

Digester

Sample