60 • Selecting the Detector Head Input Type and Gas Setup

Digester Gas Monitor Operator’s Manual

Use the UP/YES or DOWN/NO button to scroll through the units choices. When
the desired units are displayed, press and release the ENTER button. The Full
Scale Readout screen appears.

USE the UP/YES or DOWN/NO button to scroll through the available choices for
the full scale setting. When the desired full scale setting is displayed, press and
release the ENTER button.

10. The Save Settings? screen appears.

11. If you do not want to save the setup, press and release the DOWN/NO button. You will

return to the Select Channel(s) screen.

Follow the instructions from step 4 above.

If you want to save the setup, press and release the UP/YES button. The display will
indicate that the settings have been saved before asking if you want to setup more
channels.

NOTE: If you defined a custom setup by selecting the CUSTOM choice in the

PreDefined Gas Setup screen, you will be required to review the channel
parameters from the Configuration Menu. See “Viewing and Changing Channel
Parameters” on page 51 for a description of the channel parameters.

12. If you do not want to setup more channels, press and release the DOWN/NO button

and the Digester Gas Monitor will exit the Input Setup Menu and begin its warm-up
sequence.

If you want to setup more channels, press and release the UP/YES button. You will
return to the Select Channel(s) screen.

Follow the instructions from step 4 above.

CHANNEL(S): 1
Full Scale Readout
100 %LEL
UP / DN then ENT / ESC

Save Settings?

[Y / N]

Select Channel

1
UP / DN then ENT / ESC

Select Channel

1
UP / DN then ENT / ESC