RKI Instruments Digester Gas Monitor User Manual

Page 79

Advertising
background image

Digester Gas Monitor Operator’s Manual

Supported Modbus Functions • 73

Registers 40002 - 40015, Channel 1 Information

Register 40002

Table 14 below shows the information in register 40002.

Registers 40003 & 40004

The 32 bits in registers 40003 and 40004 contain the channel 1 numerical reading in
single precision floating point IEEE format.

Register 40005

The 16 bits in register 40005 contain the channel 1 numerical reading in integer form
omitting the sign (+ or -) and the decimal point.

Register 40006

Table 15 below shows the information in register 40006.

Registers 40007 - 40010

The 8 bytes (characters) in these registers contain the channel gas name in ASCII
characters.

Registers 40011 - 40013

The 6 bytes (characters) in these registers contain the channel gas reading in ASCII
characters.

Table 14: Register 40002

Channel Operational State

(MS-Byte)

Channel Status Bits

(LS-Byte)

0 = (this code is not used)

1 = Standby

2 = Normal Operation

3 = Warm Up

4 = Fail

5 = Post-Calibration

6 = Low Power

Bit-7 (msb) = Relay-1 (1 = energized)

Bit-6 = Relay-2 (1 = energized)

Bit-5 = (this bit is not used)

Bit-4 = Alarm-1 (1 = asserted)

Bit-3 = Alarm-2 (1 = asserted)

Bit-2 = Alarm-3 (1 = asserted)

Bits[1:0] = Channel Usage Code
[00] = Not Used

[01] = Standby

[10] = Active

[11] = unused code

Table 15: Register 40006

Channel Reading Sign (+/-)

(MS-Byte)

Channel Decimal Point Position

(LS-Byte)

0 = Positive (+)

1 = Negative (-)

0 = No Decimal Point

1 = .x

2 = .xx

3 = .xxx

Advertising