3.4 Setting up the needle holder

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

863 Compact Autosampler

2 Remove the nut

Loosen and remove the nut screwed onto the needle holder.

3 Insert the needle holder

Insert the needle holder in the lift head and screw it tight with the
nut from below.

4 Insert the needle

■

Loosen the pressure screw from the needle holder and remove it.

■

Insert the needle part way from above into the opening of the
needle holder.

■

Slide the PEEK ring wedge down over the needle from above. The
narrow side of the seal must face upwards.

5 Fasten the needle

■

Screw the pressure screw into the needle holder from above. At
the same time push the needle slightly upward from below to
apply a little pressure.

■

Tighten the pressure screw firmly.

6 Connect the capillary

■

Slide a 6.2744.010 PEEK pressure screw over the end of the capil-
lary.

■

Manually screw tight the PEEK pressure screw with the capillary in
the pressure screw resting on the needle holder. The capillary
must be pressed tight while doing so.

7 Close the safety shield

Tip down the safety shield and fix it with the fastening screws.

Warning

The safety shield must always be closed for safety reasons during oper-
ation of the 863 Compact Autosampler.