3 Installation

863 Compact Autosampler

■■■■■■■■

15

3.5

Installing the peristaltic pump

1

2

3

4

5

6

7

8

1

9

10

3

6.2744.160

Figure 7

Installing the pump tubing

1

PEEK pressure screws, short
(6.2744.070)

2

Tubing olive (6.2744.034)

3

Stopper
The colors of the stopper indicate the inner
diameter of the pump tubing.

4

Tubing cartridge (6.2755.000)

5

Contact pressure lever

6

Union nut

7

Adapter

8

Tubing olive

9

Pump tubing (6.1826.xx0)

10 Snap-action lever

Mount the pump tubing as follows:

1 Removing the tubing cartridge

Release the tubing cartridge from the cartridge holder by pressing
the snap-action lever and unhooking from the mounting pins (see
Figure 3, page 9)
.

2 Connecting the aspiration side

Place a 6.2744.034 tubing olive (7-2) on the aspiration side of the
pump tubing.