7.1 Automation: Sample with air gap
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
863 Compact Autosampler
7 Parameters
7.1
Automation: Sample with air gap
Menu
▶ Parameters
Automation
Display of the template used for the automation sequence.
Sample pump time
The time during which the peristaltic pump aspirates sample solution.
Range
0 ... 999 s
Default value
120 s
Air gap pump time
The time during which the peristaltic pump aspirates air for an air gap.
Range
0 ... 999 s
Default value
6 s
Pump rate
Rate of the peristaltic pump. One step corresponds to 6 rpm. Positive val-
ues = shift direction is clockwise, negative values = shift direction is coun-
terclockwise.
Range
-7 ... 7
Default value
3
7.2
Automation: Sample no air gap
Menu
▶ Parameters
Automation
Display of the template used for the automation sequence.
Sample pump time
The time during which the peristaltic pump aspirates sample solution.
Range
0 ... 999 s
Default value
120 s