4 Automation sequences

863 Compact Autosampler

■■■■■■■■

27

■

Wait for the pump time to be completed

■

Switch off the peristaltic pump

■

Wait for the stepping pulse

4.5

VA 797 Remote start

This method template is suitable for sample transfer to a 797 VA Compu-
trace. The actual pump time is determined by the control software.

The sample series is started on the 797 VA Computrace (Remote Start).

Preparing and starting the sample series

1 Press [START].

2 Under Number of samples, enter the number of samples or the

maximum value of 99.

3 Under Next sample pos., enter the rack position of the first sample

to be processed.

4 Press [BACK].

5 Start the sample series in the VA Computrace software.

Note

The method run will be automatically restarted for every sample via
remote connection (Autostart). A stepping pulse of the connected
instrument is expected at the end of the method run. The 863 Compact
Autosampler must be stopped manually by means of the button
[STOP] after the processing of the last sample.

The individual steps of the method:

■

Move to the sample

■

Lower the lift to the work position

■

Wait for the stepping pulse

■

Switch on the peristaltic pump for the sample transfer

■

Wait for the minimum pump time to be completed

■

Wait for the stepping pulse

■

Switch off the peristaltic pump

■

2 seconds waiting time

■

Wait for the stepping pulse