10 appendix, 1 remote interface, 1 pin assignment of the remote interface – Metrohm 863 Compact Autosampler User Manual
Page 70: Pin assignment of the remote interface, Figure 17, Pin assignment of remote socket and plug
Advertising

10.1 Remote interface
■■■■■■■■■■■■■■■■■■■■■■
60
■■■■■■■■
863 Compact Autosampler
10 Appendix
10.1
Remote interface
10.1.1
Pin assignment of the remote interface
1
5
6
9
5
1
9
6
Figure 17
Pin assignment of remote socket and plug
The above figure of the pin assignment applies for all Metrohm instru-
ments with 9-pin D-Sub remote connectors.
Table 3
Inputs and outputs of the remote interface
Pin No.
Assignment
Function
1
Output 0
Ready
2
Output 1
3
Output 2
4
Output 3
5
Output 4
Error
6
0 volt (GND)
7
+5 volts
8
Input 0
Start/Continue
9
Input 1
Stop
Inputs
+5 V
approx. 5 k
Ω Pull-up
tp
t
p
> 100 ms
active = low, inactive = high