4.6 VA 797 Manual start

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

863 Compact Autosampler

4.6

VA 797 Manual start

This method template is suitable for sample transfer to a 797 VA Compu-
trace. The actual pump time is determined by the control software.

Preparing and starting the sample series

1 Press [START].

2 Under Number of samples, enter the number of samples or the

maximum value of 99.

3 Under Next sample pos., enter the rack position of the first sample

to be processed.

4 Press [START].

Note

The method run will be automatically restarted for every sample via
remote connection (Autostart). A stepping pulse of the connected
instrument is expected at the beginning of the method run. The 863
Compact Autosampler may need to be stopped manually by means of
the button [STOP] after the processing of the last sample.

The individual steps of the method:

■

Wait for the stepping pulse

■

Move to the sample

■

Lower the lift to the work position

■

Wait for the stepping pulse

■

Switch on the peristaltic pump for the sample transfer

■

Wait for the minimum pump time to be completed

■

Wait for the stepping pulse

■

Switch off the peristaltic pump