Metrohm 815 Robotic Soliprep for LC User Manual
Page 40

2.16 Installing the Dosino
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
815 Robotic Soliprep for LC
Preparing the Dosino 1 for sample transfer
1
2
3
4
1 Screw an adsorber tube filled with cotton to the Vent connector on
the 10 mL dosing unit.
2 Equip the dosing unit with a filling tubing (Port 2 on the underside of
the dosing unit) and screw it onto a 6.1808.070 clear glass bottle (2 L,
with GL 45 thread). The bottle should be filled with the solvent (water
or an organic solvent) which is also used for sample dispersion.
3 Connect the transfer tubing to tower 1 with the end that is still free to
Port 1 on the dosing unit.
4 Place the Dosino (dosing drive) on the dosing unit.
5 Equip the connection cable of the Dosino with an anti-interference
adapter made of ferrite (T2400.102) (see drawing).
Connect the connection cable to the MSB connector 1 (MSB 1) on the
rear of the Sample Processor.