5 lift position for the luer connector, Lift position for the luer connector – Metrohm 815 Robotic Soliprep for LC User Manual
Page 56

5.5 Lift position for the Luer connector
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
815 Robotic Soliprep for LC
9 Under Robotic arm position, as Target position select External
1 and click on [Start].
5.5
Lift position for the Luer connector
External position 3
1 Under Device selection
▶ All devices, click on 815 Robotic
USB... and then select the tab Tower 1.
2 Under Robotic arm position actuate the button [Arrow right] until
the Luer adapter is located above the Luer connector.
3 Move the lift downwards until the Luer adapter is located approx. 1
mm above the Luer connector.
4 Under Robotic arm position actuate the arrow buttons [Arrow
left] and [Arrow right] until both components fit to one another in
the middle.
5 Click on [Configure rack].
6 Under Robotic arm position select External position 3.
7 Click on the associated [Assign] button.
8 Click on [Close].