4 Adjusting rack and robotic arm
815 Robotic Soliprep for LC
■■■■■■■■
43
Correcting the rack offset
If there is a deviation on the R Line, proceed as follows:
1 In MagIC Net, select Configuration.
2 Double-click on the device name 815 Robotic USB....
3 Switch over to the Rack tab and open the rack parameter settings with
a click on [Rack data].
4 Correct the value for Rack offset according to the observed deviation
from the positioning reticle. One tick mark corresponds to approxi-
mately 0.5°.
5 Close the rack data configuration and the Sample Processor properties
dialog each with [OK].
6 In manual control, select the same rack position again and lower the
lift down to the positioning reticle.
Now the adapter tip should point to the middle of the positioning reti-
cle. If this is not the case, then an additional correction must be made.