Metrohm 833 IC Liquid Handling Unit User Manual

Page 36

Advertising
background image

4 833 IC Liquid Handling Pump Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

28

1 Connect column to injector

•

Remove end caps from column 53.

•

without column heating:

Screw inlet end of separating column 53

(note flow direction)

to column connection capillary 50 mounted on the injector.

•

with column heating:

Prepare column heating according to Section 2.9.2 of the

819/820 Instructions for Use and screw column connection

capillary 50 (see Fig. 20, 819/820 Instructions for Use) with a

compression fitting to injection valve 51.

•

With precolumn:

Install precolumn according to the supplied leaflet between

inlet of the separating column and the injection valve.

2 Rinse column

•

Place a beaker beneath the column outlet.

•

Start 818 IC Pump in «IC Net» with suitable flow (see leaflet of

the column) and rinse column for ca. 10 min with eluent.

•

Stop 818 IC Pump.

3 Connect column to suppressor module (Suppressor

connection 1)

•

Cut inlet capillary 64 (marked with "Eluent") of suppressor

module 70 to the required length. Use the 6.2126.080

Capillary cutter available as an option.

•

without column heating

Screw inlet capillary 64 on to the outlet end of separating

column 53 using a 6.2744.010 compression fitting.

•

with column heating

Connect capillary 52 at the outlet of the separating column

(see Fig. 20, 819/820 Instructions for Use) to the inlet

capillary 64 using the 6.2744.040 Coupling.

4 Fix column

•

without column heating

Insert one or two column holders 59 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 58 and fasten separating

column in the column holder.

•

with column heating

Insert column heating according to Fig. 10 (819/820

Instructions for Use) into the Separation Center.