5.1 Connection of the Suppressor module

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

33

5 Connect column to suppressor module (Suppressor

connection 1)

•

Cut inlet capillary 27 (marked with “Eluent”) at suppressor

connection 25 (see Fig. 9) to the required length. Use the

6.2126.080 Capillary cutter available as an option.

•

without column heating

Screw inlet capillary 27 on to the outlet end of separating

column 53 using a 6.2744.010 compression fitting.

•

with column heating

Connect capillary 52 at the outlet of the separating column

(see Fig. 20, 819/820 Instructions for Use) to the inlet

capillary 27 using the 6.2744.040 Coupling supplied with the

820 standard equipment.

6 Fix column

•

without column heating

Insert one or two column holders 59 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 58 and fasten separating

column in the column holder.

•

with column heating

Insert column heating according to Fig. 10 (819/820

Instructions for Use) into the Separation Center.

7 Connect suppressor module to detector block

(Suppressor connection 1)

•

Cut outlet capillary 32 (marked with “Detector”) at suppressor

connection 25 (see Fig. 9) to coupling 71(6.2620.060) to the

required length. Use the 6.2126.080 Capillary cutter available

as an option.

•

Screw outlet capillary 32 on to coupling 71 by using a

6.2744.010 compression fitting.

•

Screw inlet capillary 56 of detector block 57 on to the other

end of coupling 71.

8 Attach pump tubing to 833 Liquid Handling Unit

•

Mount both the 6.1826.050 Pump tubing (white-yellow

stopper) as described in Section 3.2, see Fig. 7.

9 Suppressor connection 2: H

2

SO

4

•

Pull inlet capillary 28 (marked with “H

2

SO

4

”) at suppressor

connection 25 (see Fig. 9) through one of the openings 18 on

the rear panel of the 820 IC Separation Center.

•

Attach inlet capillary 28 (23 in Fig. 7) to the filter unit PEEK 22

(6.2821.120) of the first pump tubing using a compression

fitting 17 (6.2744.010), see Fig. 7.

•

Mount aspiration tubing: cut a piece of PTFE tubing 16

(6.1803.020) to the required length.

•

Attach a PEEK compression fitting 17 (6.2744.010) to one

end of the PTFE tubing 16 and screw this on to coupling 18.