The 833 ic liquid handling sample preparation unit – Metrohm 833 IC Liquid Handling Unit User Manual

Page 47

Advertising
background image

6.1 Connection to modular IC system for neutralization

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

39


36 Inlet capillary for sample

37 Inlet capillary for acid

38 Outlet capillary for acid

39 Outlet capillary for H

2

O

40 Inlet capillary for H

2

O

41 Outlet capillary for sample


Fig. 13:

Connections at sample preparation module of the 833 IC Liquid
Handling Sample Preparation Unit

1 Insert sample preparation module

•

Place sample preparation module 33 inside the chamber on

the floor.

2 Connect sample preparation module

•

Remove plastic stoppers from the rear panel openings 18 of

the 820 IC Separation Center and lead cable 35 mounted on

sample preparation module 33 back through this opening.

•

Connect cable 35 to connection 13 “Module” of the 833 IC

Liquid Handling Unit (see Fig. 2).

3 Connect sample feed to sample preparation

module (Sample Preparation connection 1)

•

Connect inlet capillary 36 (marked with “Sample in”) at the

sample preparation module connection 34 (see Fig. 13) to

connection “5” of injection valve B using a 6.2744.010

compression fitting.

4 Connect sample preparation module to valve A

(enrichment column) (Sample Preparation

connection 1)

•

Connect outlet capillary (marked “Sample out”) at sample

preparation module connection 34 (see Fig. 13) to

connection “1” of injection valve A using a 6.2744.010

compression fitting.

•

When making the further connections to injection valve A

make sure that the ”Inject” setting of the enrichment column

is rinsed in the reverse direction to the “Fill” setting.

Sample in

H

2

O

Sample out

Waste

Waste

Regenerant

1

3

2

39

36

37

38

40

41