7 833 IC Liquid Handling Dialysis Unit

833 IC Liquid Handling Unit / Instructions for Use 8.833.1003

66

7.7 Dialysis

procedure

For determinations of samples by using dialysis it is advisable to

proceed in the following sequence:

1 Prepare acceptor solution

•

To avoid interference from air bubbles in the acceptor

channel the ultrapure water used for the acceptor solution

should always be degassed for at least 10 min by vacuum,

N

2

or He.

2 Prepare sample

•

In order to avoid blocking the sample channel, samples

containing a large amount of suspended or solid particles

should always be centrifuged using a benchtop centrifuge

at 10'000 min

-1

for 5 min.

3 Start up IC system

•

Switch on all instruments.

•

Start the IC Net software and load/create the required system

(see «IC Net» Instructions for Use).

•

Condition the IC column (see 819/820 Instructions for Use).

4 Start up dialysis system

•

Immerse both pieces of aspiration tubing 16 in the acceptor

solution.

•

Start the peristaltic pumps in «IC Net» by setting the remote

line 1 (833 Pump Unit) and remote line 2 (833 Dialysis Unit) in

the window of the 819 Detector.

•

Rinse the dialysis system with acceptor solution for approx.

10 min and then switch off the pump again.

5 Calibration

•

Immerse sample aspiration tubing 16 in the standard

solution.

•

Start the determination in IC Net and wait until the standard

chromatogram has been evaluated and the next standard

can be measured.

6 Sample determination

•

Immerse sample aspiration tubing 16 in the sample solution.

•

Start the determination in IC Net and wait until the

chromatogram has been evaluated.